Title of article :
A simple solution to systematic errors in density determination
by X-ray reflectivity: The XRR-density evaluation
(XRR-DE) method
Author/Authors :
P. Bergese *، نويسنده , , E. Bontempi، نويسنده , , L.E. Depero، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
X-ray reflectivity (XRR) is a non-destructive, accurate and fast technique for evaluating film density. Indeed, sample-goniometer alignment is a
critical experimental factor and the overriding error source in XRR density determination. With commercial single-wavelength X-ray
reflectometers, alignment is difficult to control and strongly depends on the operator. In the present work, the contribution of misalignment
on density evaluation error is discussed, and a novel procedure (named XRR-density evaluation or XRR-DE method) to minimize the problem will
be presented. The method allows to overcome the alignment step through the extrapolation of the correct density value from appropriate nonspecular
XRR data sets. This procedure is operator independent and suitable for commercial single-wavelength X-ray reflectometers. To test the
XRR-DE method, single crystals of TiO2 and SrTiO3 were used. In both cases the determined densities differed from the nominal ones less than
5.5%. Thus, the XRR-DE method can be successfully applied to evaluate the density of thin films for which only optical reflectivity is today used.
The advantage is that this method can be considered thickness independent
Keywords :
density , X-ray reflectivity (XRR) , Sample-goniometer , XRR-density evaluation (XRR-DE) , Thin films
Journal title :
Applied Surface Science
Journal title :
Applied Surface Science