Title of article :
Three-gamma annihilation imaging in positron emission tomography
Author/Authors :
K.، Kacperski, نويسنده , , N.M.، Spyrou, نويسنده , , F.A.، Smith, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
It is argued that positron annihilation into three photons, although quite rare, could still be used as a new imaging modality of positron emission tomography. The information gained when the three decay photons are detected is significantly higher than in the case of 511 keV two-gamma annihilation. The performance of three-gamma imaging in terms of the required detector properties, spatial resolution and counting rates is discussed. A simple proof-of-principle experiment confirms the feasibility of the new imaging method.
Journal title :
IEEE Transactions on Medical Imaging
Journal title :
IEEE Transactions on Medical Imaging