• Title of article

    Evaluation of nanoscale roughness measurements on a plasma treated SU-8 polymer surface by atomic force microscopy

  • Author/Authors

    Ferdinand Walther، نويسنده , , Wolfgang M. Heckl، نويسنده , , Robert W. Stark، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    6
  • From page
    7290
  • To page
    7295
  • Abstract
    A comparison between roughness data obtained with an atomic force microscope (AFM) on different surfaces requires reliable roughness parameters. In order to specify the appropriate parameters for nanoscale roughness measurements, we compared the root mean square (rms) roughness and the relative surface area (sdr) as function of varying scan size, speed and pixel size. By using oxygen plasma (24 kJ) treated SU-8 with an average rms roughness of 2.6 ± 0.5 nm as reference surface, the repeatability of the method was evaluated for dynamic (tapping) and contact mode. The evaluation of AFM images indicated a decrease of the effective tip radius after a few measurements. This degradation of the tip lowers the resolution of the image and can affect roughness measurements.
  • Keywords
    Hydrophilisation , Intermittent contact , AFM , Tip degradation , SU-8 , Photoresist , Plasma treatment , Polymer
  • Journal title
    Applied Surface Science
  • Serial Year
    2008
  • Journal title
    Applied Surface Science
  • Record number

    1009692