Title of article :
Continuous flow-injection-atomic absorption spectrometric method for the determination of Ondansetron
Author/Authors :
L.Lahuerta Zamora، نويسنده , , J.Mart?nez Calatayud، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
6
From page :
143
To page :
148
Abstract :
A flow-injection procedure for the indirect determination of the new drug Ondansetron is proposed. The method is based on the reaction of the drug in an oxidative solid-phase reactor included in the flow assembly. The reactor was made by lead dioxide physically entrapped by polymerization; the released lead(II) was monitored by atomic absorption spectrometry at 217.0 nm. The procedure gave a linear calibration graph up to 20 μg ml−1 of Ondansetron with a sample throughput of 338 samples h−1.
Keywords :
Atomic absorption spectrometry , Solid-phase reactors , Pharmaceutical analysis , Flow injection , Ondansetron
Journal title :
Analytica Chimica Acta
Serial Year :
1995
Journal title :
Analytica Chimica Acta
Record number :
1022181
Link To Document :
بازگشت