• Title of article

    Comparison of two methods for coating piezoelectric crystals

  • Author/Authors

    Teresa Gomes، نويسنده , , Armando C. Duarte، نويسنده , , Jo?o P. Oliveira، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    6
  • From page
    329
  • To page
    334
  • Abstract
    Quartz piezoelectric crystals were coated with triethanolamine using two procedures: spray and syringe methods. The reproducibility of both coating methods was evaluated comparing the relative standard deviation of batches of four crystals coated by each of the procedures. The relative standard deviation obtained with the spray method is a hyperbolic function of the coating frequency and is, at least, three times lower than the one calculated from results of the syringe method, which is not frequency dependent. An explanation for the poorer reproducibility of the syringe method is given, based on the radial sensitivity theory and the non-uniformity of the coating distribution over the crystal, as evidenced by microscopic photographs.
  • Keywords
    Piezoelectric crystals , Quartz Crystal Microbalance , Coatings
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1995
  • Journal title
    Analytica Chimica Acta
  • Record number