Long-term response of an integrated carbon paste based glucose biosensor
Author/Authors
J. Kulys، نويسنده , , H.E. Hansen، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
10
From page
285
To page
294
Abstract
A biosensor containing integrated reference, auxiliary and two carbon paste based bioelectrodes shows linear calibration graphs for glucose concentrations > 16 mM. The biosensor response is performed on a renovated surface at an electrode potential of 60–80 mV; the capacity of the biosensor is 100 measurements. The long-term response of the biosensor relies on a glucose oxidase stability that depends on the structure of oil, water activity and stabilizer. The response of the biosensors containing silicone oil and stabilized glucose oxidase hardly changes during a period of three months at 35 °C and a relative humidity of 75%.
Keywords
Biosensors , Carbon paste , Glucose sensor