Title of article :
Construction and use of a tubular picrate ion-selective electrode for reducing sugar determination in Port wine by flow-injection analysis
Author/Authors :
Teresa I.M.S. Lopes، نويسنده , , Antonio O.S.S. Rangel، نويسنده , , JoséL.F.C. Lima، نويسنده , , M.Conceiç?o B.S.M. Montenegro، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
A picrate ion-selective electrode without inner reference solution and based on bis(triphenylphosphoranylidene)ammonium picrate dissolved in 2-nitrophenyloctyl ether was constructed and its characteristics assessed. With respect to reducing sugar determination in Port wine, a tubular electrode based on the same sensor system was developed and incorporated into a flow-injection manifold devised for this determination. The methodology involves a reaction between the reducing sugar and picric acid, the decrease in the picrate concentration being monitored with the tubular electrode. A linear dependence between peak width, at a fixed potential and the logarithm of sugar concentration in the range 25–200 g l−1 was obtained. The proposed system allows wine samples to be injected with no previous treatment and allows a sample throughput of 50 determinations per hour, a relative standard deviation less than 4%, and the results are comparable to the reference procedures.
Keywords :
Flow injection , Ion selective electrodes , Wine , Sugar , Picrate
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta