• Title of article

    Flow-injection spectrofluorimetric determination of sulphate using calcein and zirconium

  • Author/Authors

    N. Chimpalee، نويسنده , , D. Chimpalee، نويسنده , , S. Suparuknari، نويسنده , , B. Boonyanitchayakul، نويسنده , , D.Thorburn Burns، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1994
  • Pages
    4
  • From page
    401
  • To page
    404
  • Abstract
    Sulphate can be determined spectrofluorimetrically (λem = 505 nm, λex = 410 nm) based on the formation of a ternary complex between sulphate, zirconium and calcein in a flow-injection system. The carrier stream was deionised water and the reagent streams were 100 μg ml−1 Zr (as ZrOCl2) in 1% (v/v) hydrochloric acid, and 40 μg ml−1 calcein in 0.1 M potassium hydroxide. The injection rate was 50 h−1. The calibration graph was linear up to 25 μg ml−1 SO2−4 based on 250-μl injection volumes. The relative standard deviation (n = 10) was 1.95% for 10 μg ml−1 SO2−4; the detection limit was 0.82 μg ml−1 SO2−4. The system has been applied to the determination of sulphur in NIST coal samples after oxygen bomb combustion.
  • Keywords
    Flow injection , Calcein , Sulphate , Zirconium , Fluorimetry
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1994
  • Journal title
    Analytica Chimica Acta
  • Record number