Title of article :
Microbiosensor based on an integrated thermopile
Author/Authors :
Bin Xie، نويسنده , , Michael Mecklenburg، نويسنده , , Bengt Danielsson، نويسنده , , Ove ?hman، نويسنده , , Fredrik Winquist، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1994
Abstract :
A microbiosensor based on an integrated thermopile was designed and fabricated on a quartz chip. The thermopile, which was manufactured by doping boron in polysilicon together with aluminium, provided a potential output of ca. 2 mVK. A silicone rubber membrane was used to form and seal the microchannel. The total column volume was 20 μl. Glucose oxidase and catalase were co-immobilized on spherical CPG beads (controlled pore glass) and in turn charged into the microchannel. Using 1 μl sample volumes, a linear range of 2 to 25 mM glucose was obtained using a flow rate of 105 μlmin. The relative standard deviation for 100 glucose samples (10 mM) was 5%.
Keywords :
Biosensors , Glucose , Micromachining , Integrated thermopile , Flow injection , Thermal biosensor
Journal title :
Analytica Chimica Acta
Journal title :