Title of article :
Analytical application of membranes with covalently bound glucose oxidase
Author/Authors :
E.G. Entcheva، نويسنده , , L.K. Yotova، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1994
Abstract :
Biosensor membranes with glucose oxidase, covalently bound to acetylcellulose carriers activated with urea, were prepared and tested in order to evaluate their application in measuring the glucose concentration in blood serum and plasma. A computer-interfaced measurement and research system was used to obtain data about the characteristics of the constructed biosensors. Stability (less than 7.5% change in sensitivity for 20 days by overnight storage at 25°C), biocompatibility (2.2–2.4% loss of sensitivity after 10 blood samples) and acceptable correlation with conventional methods were found to be the main advantages of the biosensors thus obtained.
Keywords :
Glucose oxidase , Biosensors , Biosensor membranes