Title of article :
Predictive model for coulometric operation in a thin-layer amperometric flow cell
Author/Authors :
Danut-Ionel Vaireanu، نويسنده , , Nigel Ruck، نويسنده , , Peter R. Fielden، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
A theoretical model is presented that may be usefully employed to establish the limiting geometric and hydrodynamic conditions for a thin-layer cell to attain 100% coulometric efficiency. The thin-layer cell, based on a glassy-carbon electrode 5 cm in length (the flow path) and 2 mm wide was tested for three different channel heights at high and low flow rates (1 and 60 ml/h) using 1 mM potassium hexacyanoferrate(II) as the electrochemical probe in conjunction with a flow injection system.
Keywords :
Coulometry , Flow injection , Thin-layer cell/detector