• Title of article

    Iodide buffer systems for the calibration of iodide-selective electrodes

  • Author/Authors

    L.P. Loginova، نويسنده , , V.D. Bazilyanskaya، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    5
  • From page
    37
  • To page
    41
  • Abstract
    A new approach to the creation of anion buffer systems is proposed, which extends the linear range. The new iodide buffer systems, containing (1) HgI2 precipitate, CaEDTA2−, Ca2+, HgEDTA2−, and (2) HgI2 precipitate, chloride complexes of Hg2+ and excess of Cl− ions, are proposed for the calibration of iodide-selective electrodes at pI > 5. It is shown with the help of these buffers, that a solid-state Orion 94-53 iodide-selective electrode keeps linear Nernstian response up to pI = 7.5. Iodide buffer systems are used for the estimation of the reliability of literature data on selectivity coefficients of iodide-selective electrodes with respect to Cl− ions.
  • Keywords
    Iodide , Ion selective electrodes
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1995
  • Journal title
    Analytica Chimica Acta
  • Record number

    1022678