Title of article :
Determination of selenium compounds in urine samples by liquid chromatography-inductively coupled plasma mass spectrometry with an ultrasonic nebulizer
Author/Authors :
Kuei-Lin Yang، نويسنده , , Shiuh-Jen Jiang، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
7
From page :
109
To page :
115
Abstract :
An inductively coupled plasma mass spectrometer (ICP-MS) coupled with ultrasonic nebulization was used as a liquid chromatographic detector for the determination of selenium compounds. Selenite, selenate, and trimethylselenonium were separated by ion-pair reversed-phase liquid chromatography before introduction into the inductively coupled plasma. Detection limits were in the range of 22–74 pg selenium for various selenium species. The relative standard deviation of the peak heights for five injections of 10 ng ml−1 selenium mixture was in the range of 2–5%. This method has been applied to determine various selenium species in urine samples.
Keywords :
Inductively coupled plasma-mass spectrometry , Liquid chromatography , Selenium speciation , Urine samples , Ultrasonic nebulization
Journal title :
Analytica Chimica Acta
Serial Year :
1995
Journal title :
Analytica Chimica Acta
Record number :
1022688
Link To Document :
بازگشت