Title of article
Mass spectrometric analysis of additives in polymer extracts by desorption chemical ionization and collisional induced dissociation with B/E linked scanning
Author/Authors
C.G. Juo، نويسنده , , S.W. Chen، نويسنده , , G.R. Her، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
12
From page
153
To page
164
Abstract
Desorption chemical ionization in combination with collisional induced dissociation and B/E linked scan proved to be a viable method for identifying additives in polyethylene extracts by mass spectrometry. In comparison with fast atom bombardment (FAB), much less fragmentation was observed with desorption chemical ionization (DCI) using ammonia as reagent gas. Furthermore, most likely due to the “matrix effect”, not all the additives in the extracts were detected by FAB. The softness and the lack of matrix effect make ammonia DCI a better ionization technique than FAB for the analysis of additives directly from the extracts. Most likely due to higher collision energy, product ion mass spectra acquired with a double focusing mass spectrometer provided more structural information than the spectra obtained with a triple quadrupole mass spectrometer. Besides contamination from the seals of the supercritical fluid extractor, high and low density polyethylenes extracted with supercritical carbon dioxide produced similar results as with toluene.
Keywords
Supercritical fluid extraction , mass spectrometry , Desorption chemical ionization , Collisional induced dissociation , Polyethylene , Polymers
Journal title
Analytica Chimica Acta
Serial Year
1995
Journal title
Analytica Chimica Acta
Record number
1022763