• Title of article

    Automated trace enrichment and determination of metals using a combination of supported liquid membrane for sample pretreatment and graphite furnace atomic absorption spectrometry for the determination

  • Author/Authors

    F. Malcus، نويسنده , , Nii-Kotey Djane، نويسنده , , L. Mathiasson، نويسنده , , G. Johansson، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    6
  • From page
    295
  • To page
    300
  • Abstract
    A computer controlled system has been developed for automated trace analysis for metals. The system used consists of four parallel supported liquid membrane devices for analyte enrichment, leaving an efficiently cleaned and enriched sample in the acceptor for the final analytical step, which is performed by graphite furnace atomic absorption spectrometry (AAS). The system was tested with 40% di(2-ethylhexyl)phosphate in kerosene as the liquid in the porous membrane and with the model metals Al, Cd and Cu at concentrations below μg/l. The average relative standard deviation (RSD) was 6.5% at 0.5 μg/l and 6.3% at 1.0 μg/l based on 108 independent measurements. The blank for Al, which is the most difficult metal to determine with respect to contamination was 0.15 μg/l with an average RSD (n = 16) of 11%. Using a 30 min period for processing the samples in the supported liquid membrane step, the sample throughput was eight samples per hour.
  • Keywords
    Atomic absorption spectrometry , Trace metals , Supported liquid membranes , Automatic on-line preconcentration
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1996
  • Journal title
    Analytica Chimica Acta
  • Record number

    1023088