Title of article :
The automation of immun-FIA-systems
Author/Authors :
B. Hitzmann، نويسنده , , A. L?hn، نويسنده , , M. Reinecke، نويسنده , , B. Schulze، نويسنده , , T. Scheper، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
In this contribution an automation system for the control, measurement uptake, evaluation and visualization is presented as well as its application for two different flow-injection immunoassays. Both systems are based on the principles of flow-injection analysis and were developed for a reliable, fast monitoring of relevant proteins in animal cell cultivation processes. One system uses a homogeneous assay with turbidimetric detection, the other a heterogeneous assay with an immobilized immunocomponent applying fluorescence detection. The on-line application during a mammalian cell cultivation as well as off-line measurements of medium samples are presented. The results obtained from on-line measurements are compared to ELISA results. The automation of the analysis systems with respect to its flexible control as well as the reliable evaluation of the measurement signals -even if they are severely distorted -are discussed in detail.
Keywords :
Flow injection , Bioprocess monitoring , Reliable evaluation , Automation , Immunoassay
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta