Title of article
Glucose-sensitive ion-sensitive field-effect transistor-based biosensor with additional positively charged membrane. Dynamic range extension and reduction of buffer concentration influence on the sensor response
Author/Authors
Vjacheslav Volotovsky، نويسنده , , Alexey P. Soldatkin، نويسنده , , Alexander A. Shulʹga، نويسنده , , Victor K. Rossokhaty، نويسنده , , Vitaliy I. Strikha، نويسنده , , Anna V. Elʹskaya، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
5
From page
77
To page
81
Abstract
Glucose-sensitive sensors based on ion-sensitive field-effect transistors were prepared by cross-linking a mixture of glucose oxidase and bovine serum albumin in saturated glutaraldehyde vapour on the sensitive area of the transducer. After this procedure, 4-vinylpyridine-styrene copolymer membranes were deposited on top of the sensors by drop-coating of 0.5% polymer solution in ethanol. The use of this additional membrane resulted in substantial extension of the dynamic range of the sensor (up to 20 mM of glucose) and reduction of buffer concentration dependence of the sensor response. Also the effects of pH and ionic strength were examined for the sensors with and without additional membranes.
Keywords
Membranes , Buffer capacity , ISFETs , Enzymatic methods , Glucose , Dynamic range , Biosensors
Journal title
Analytica Chimica Acta
Serial Year
1996
Journal title
Analytica Chimica Acta
Record number
1023973
Link To Document