• Title of article

    Determination of low levels of platinum group metals: new solutions

  • Author/Authors

    I.V. Kubrakova، نويسنده , , T.F. Kudinova، نويسنده , , N.M. Ku?min، نويسنده , , I.A. Kovalev، نويسنده , , G.I. Tsysin، نويسنده , , Yu.A. Zolotov، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    9
  • From page
    167
  • To page
    175
  • Abstract
    A new rapid sample preparation scheme for the determination of low levels of platinum metals (PMs) has been proposed. The technique includes, microwave dissolution of samples followed by dynamic sorption preconcentration of the PMs on a polystyrene based sorbent containing diethylenetriamine groups. Non-equilibrium dynamic conditions allowed simultaneously and nearly quantitative recovery of all the PMs in the presence of a 106–107 excess of matrix elements. Metals were determined in slurries of concentrates by Zeeman electrothermal atomic absorption spectrometry. A good agreement between the results obtained and the certified values was demonstrated. The relative standard deviation is generally 0.05. Detection limits obtained were 10−6–10−7%.
  • Keywords
    Microwave decomposition , Platinum metals , Dynamic sorption preconcentration , Atomic absorption spectrometry
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1996
  • Journal title
    Analytica Chimica Acta
  • Record number

    1024309