• Title of article

    Direct vapor generation Fourier transform infrared spectrometric determination of ethanol in blood

  • Author/Authors

    Amparo Pérez-Ponce، نويسنده , , Salvador Garrigues، نويسنده , , Miguel de la Guardia، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    7
  • From page
    123
  • To page
    129
  • Abstract
    A new procedure is proposed for a direct determination of ethanol in plasma and whole blood. The method is based on the injection of a discrete sample volume of 10 μl into an electrically heated Pyrex glass reactor in which, at a temperature of 90 °C, the ethanol is volatilized and introduced by means of a N2 carrier flow inside a long-path infrared gas cell and the corresponding flow analysis recording registered as a function of time. The measurement of the area of the flow injection recording, obtained from the absorbance of the transient signal in the range 1150−950 cm−1, allows the direct quantification of ethanol upto 2 g l−1, with a limit of detection of 0.020 g l−1 and coefficient of variation between 0.3 and 1.9% for three replicate analyses of the same sample. The analysis frequency of the method is 40 h−1, and it can be applied to a single drop of finger blood.
  • Keywords
    Row injection , Vapor generation , Blood , Infrared spectrometry , Fourier transform , Ethanol determination
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1996
  • Journal title
    Analytica Chimica Acta
  • Record number

    1024350