• Title of article

    Simultaneous immunoassay using piezoelectric immunosensor array and robust method

  • Author/Authors

    Xia Chu، نويسنده , , Jianhui Jiang، نويسنده , , Guoli Shen، نويسنده , , Ru-Qin Yu، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    9
  • From page
    185
  • To page
    193
  • Abstract
    A simultaneous immunoassay technique has been developed for the determination of dual analytes, immunoglobulin M (IgM) and C-reactive protein (CRP) by constructing a piezoelectric quartz crystal array system. This array system was composed of five piezoelectric quartz crystals. Each crystal was immobilized with an antibody mixture, which had a fixed ratio of anti-IgM to anti-CRP antibodies but worked under different detection conditions, i.e., in five buffer media with different pH values. The specificity of immunoreactions and the effect of pH on the sensitivity of the immunosensors were investigated. Applications of this array system in the estimation of the concentration of unknown samples using three linear regression methods, i.e., ordinary least squares (OLS), partial least squares (PLS) and least terminated squares (LTS), were presented. Results of these samples showed that the LTS estimator is a promising tool for multi-analyte piezoelectric immunoassay.
  • Keywords
    LTS , Piezoelectric quartz crystal array , Immunoassay , Robust regression , PLS