Title of article :
Studies on a potentiometric urea biosensor based on an ammonia electrode and urease, immobilized on a γ-aluminum oxide matrix
Author/Authors :
Baohong Liu، نويسنده , , Renqi Hu، نويسنده , , Jiaqi Deng، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
9
From page :
161
To page :
169
Abstract :
A potentiometric biosensor for urea is developed, based on the immobilization of urease on a γ-Al2O3 substrate without using bovine serum albumin-glutaraldehyde as usual. The large specific surface area and porous structure of the γ-Al2O3 particle result in high enzyme loading, and the urease entrapped in this matrix is stable. FT-IR spectra were employed to give insight into the intermolecular interaction between the substrate and urease. The point of zero charge calculated from mass titration was used to explain the influence of pH on immobilization. The developed biosensor is sensitive to urea in the linear range 3.0 × 10−5-1.4 × 10−2 M, with a detection limit of 10 μM. Some studies on γ-Al2O3-urease were reported and the characteristics of the sensor were examined. The novel urea sensor was also successfully applied for the actual measurement of urea in urine.
Keywords :
Biosensors , ?-Aluminum oxide (?-Al2O3) , Urease , Isoelectric point
Journal title :
Analytica Chimica Acta
Serial Year :
1997
Journal title :
Analytica Chimica Acta
Record number :
1024449
Link To Document :
بازگشت