  • Abstract
    An amperometric biosensor for salicylate detection was developed by immobilizing salicylate hydroxylase via glutaraldehyde onto a polypyrrole film doped with hexacyanoferrate, supported on a glassy carbon electrode surface. The sensor monitors the catechol produced in the enzymatic reaction on the film surface, at an applied potential of 150 mV vs. SCE. A [NADH]/[salicylate] ratio between 2 and 4 gave the best response. The biosensor presented the best performance in a solution with pH = 7.4. The response time was about 40 s. A linear range of response was observed for salicylate concentrations between 1.0 × 10−5 and 1.0 × 10−4mol l−1 and the equation adjusted for this curve was I = (−0.04 ± 0.01) + (11.4 ± 0.2)[salicylate] with a correlation coefficient of 0.999 for n = 6. The biosensor retains its activity for at least 10 days despite daily use. The results obtained using the biosensor for salicylate determination, in three different samples of antithermic drugs, presented a good correlation with the standard colorimetric method.
  • Keywords
    Salicylate biosensor , Polypyrrole film , Amperometry , Salicylate hydroxylase