Title of article
Detection, interpretation and correction of changes in the instrumental response of near-infrared monochromator instruments over time
Author/Authors
E. Bouveresse، نويسنده , , S.C. Rutan، نويسنده , , Y. Vander Heyden، نويسنده , , W. Penninckx، نويسنده , , D.L. Massart b، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
19
From page
283
To page
301
Abstract
A new approach is proposed for the detection of changes in the instrumental response of near-infrared spectrometers over time. This approach is based on simulating possible instrumental differences and performing a principal component analysis of the simulated data set. This multivariate approach is tested on a real near-infrared data set and the results are compared to those obtained by the several univariate diagnostic procedures commonly used to check the stability of the response of a nearinfrared spectrometer over time. Advantages and drawbacks of this new approach are discussed. A simulated data set is used to study the influence of using different types of standards on the results obtained by this approach.
Journal title
Analytica Chimica Acta
Serial Year
1997
Journal title
Analytica Chimica Acta
Record number
1024603
Link To Document