• Title of article

    Automated pH-ISFET measurements under hydrostatic pressure for marine monitoring application

  • Author/Authors

    N Le Bris، نويسنده , , D Birot، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    11
  • From page
    205
  • To page
    215
  • Abstract
    Association of microsensors and flow measurement techniques could provide many advantages for the in-situ monitoring of marine environment. First, adaptability of such sensors to autonomous measurement and under pressure conditions has to be considered. This paper presents a laboratory study devoted to seawater pH flow measurement using ISFETs. The measurement principle had to circumvent drift problems associated with these sensors. This was performed by measuring potential steps produced by a fast alternation of the sample and a standard buffer over the sensor. In addition, interference of seawater ionic matrix on ISFET response was circumvented by the use of saline buffers. The precision of the method is better than 0.01 pH and could be further increased. Its optimal accuracy will be limited to 0.01 pH, due to uncertainties on the pH of saline buffers. The ability of ISFETs to operate under pressure, up to 30 bars, over 43 h was demonstrated, without significant decrease in sensitivity. The method could also be suitable for long-term unmanned measurements. These first results incline to further examination of the potentialities of this technique for pH monitoring at depth.
  • Keywords
    Marine monitoring , ISFET , Flow measurements , Underwater measurement , Seawater pH , pH sensor
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1997
  • Journal title
    Analytica Chimica Acta
  • Record number

    1024889