Title of article
Determination of free and total fluoride in rain water using a continuous-flow system equipped with a fluoride ion-selective electrode detector
Author/Authors
Hirokazu Hara، نويسنده , , Keiko Yabuuchi، نويسنده , , Mayumi Higashida، نويسنده , , Masahiro Ogawa، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
7
From page
117
To page
123
Abstract
A continuous-flow system equipped with a fluoride ion-selective electrode (F− ISE) detector was constructed for the determination of free and total amounts of fluoride in rain water. The buffer system proposed was a mixture of β-alanine (2-aminopropionic acid) and HCl with and without CDTA (1,2-cyclohexanediamine tetraacetic acid). Only one calibration graph is needed to calculate the free and total fluoride concentrations. The detection limit is 1×10−8 mol dm−3. The precision of five independent measurements was within 7.3% for 3×10−8 – 1×10−6 mol dm−3 standard fluoride solutions. About 50–70% of the fluoride in rain water samples was found to be in free form.
Keywords
Fluoride ion-selective electrode , Continuous-flow analysis , Speciation of fluoride , Rain water analysis
Journal title
Analytica Chimica Acta
Serial Year
1998
Journal title
Analytica Chimica Acta
Record number
1025045