• Title of article

    Application of molecular emission cavity analysis to the determination of tin in various samples based on hydride generation

  • Author/Authors

    Ali Celik، نويسنده , , Emür Henden، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    6
  • From page
    295
  • To page
    300
  • Abstract
    Determination of the total amount of tin by molecular emission cavity analysis (MECA) after generation of stannane by means of reduction with sodium tetrahydroborate(III) and measuring the SnH emission within the flame-containing MECA cavity is described. The volatile hydrides of tin, germanium and arsenic were separated on a short gas chromatographic column. The calibration graph was linear over the range 5–250 ng Sn, and the detection limit (SN = 21) was 5 ng Sn, and the relative standard deviation for the determination of 50 ng Sn was 2.8% (n = 8). The method was applied successfully to the determination of total tin in fish, canned soft drinks and sea sediment samples. Hydride generation recovery from soil sample solutions, however, was low.
  • Keywords
    Hydride generation , MECA application , TIN
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1996
  • Journal title
    Analytica Chimica Acta
  • Record number