Title of article :
The 2.5th order differential voltammetric determination of phenol with a composite carbon paste/polyamide electrode
Author/Authors :
Yongde Zou، نويسنده , , Jinyuan Mo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Abstract :
The 2.5th order differential voltammetric behavior of phenol at a composite carbon paste/polyamide electrode in aqueous solution is reported. Adsorption of the phenol onto the electrode under open-circuit conditions is followed by application of the 2.5th order differential staircase sweep voltammetry. The results obtained under optimum conditions allowed the development of a method to determine phenol in the ranges of 1 × 10−5 to 6 × 10−5 mol 1−1, 1 × 10−6 to 1 × 10−5 mol 1−1, 1 × 10−7 to 1 × 10−6 mol 1−1 and 5 × 10−8 to 1 × 10−7 mol 1−1. The relative standard deviation is 2.2% for a concentration level of 5.6 × 10−6 mol 1−1and detection limit of 8.5 × 10−9mol 1−1 (RSD = 4.5%) was achieved. The effect of experimental conditions and the electrode mechanism were investigated. Good results were obtained by applying the proposed method to the determination of phenol in cola drinks.
Keywords :
Carbon paste electrode , Phenol , Polyamide 2.5th Order differential voltammetry
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta