• Title of article

    Determination of hydrogen peroxide using a potassium hexacyanoferrate(III) modified titanium dioxide electrode

  • Author/Authors

    Yuji Mishima، نويسنده , , Junko Motonaka، نويسنده , , Kenichi Maruyama، نويسنده , , Sanae Ikeda، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    6
  • From page
    291
  • To page
    296
  • Abstract
    A titanium dioxide electrode modified with potassium hexacyanoferrate(III) was developed. Since potassium hexacyanoferrate(III) has a rapid electron transfer rate, the conductivity of the modified electrode was much better than at a bare titanium dioxide electrode. With this electrode, hydrogen peroxide (from 1.66×10−6 to 3.22×10−3 M) could be determined accurately with a relative standard deviation of less than 7.0%, in the presence of a molar ratio of 1:10 sodium chloride, potassium chloride, potassium bromide, sodium sulfate, β-d-glucose, or nicotinamide, 1:1 sodium iodide, sulfonic acid, nitric acid, uric acid, or l-ascorbic acid or 1:0.1 l-serine or l-phenylalanine.
  • Keywords
    Hydrogen peroxide , Potassium hexacyanoferrate(III) , Titanium dioxide
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1998
  • Journal title
    Analytica Chimica Acta
  • Record number

    1025463