Abstract
The construction and general performance characteristics of potentiometric sensors
responsive to some anti-histaminics Hydroxyzine hydrochloride (HYZ), Meclozine
hydrochloride (MOZ) were described. The membranes incorporate the ion-association
complex of the drugs with Sodium 12-Tungstophosphate (TP) sensor-1, Ammonium
Reineckate (AR) senser-2, and Ammonium Molbydate(AM) senser-3as electroactive
materials. These sensors exhibit fast, stable, and near-Nernstain response over the
concentration range 10-2 to 10-7 M with slopes of 53.07 and 52.14 mV per concentration
decade for sensors 1, 2. No interferences are caused by many inorganic and organic cations.
The sensors have a fast response time (20-40 s) and applicaple over a wide range of pH (2-7),
low detection limit (1.8×10-7 M) and good stability (5-6 weeks).The sensors were used for
direct potentiometric of anti-histaminics in some pharmaceutical preparations. Validation of
the method according to the quality assurance standards shows suitability of the proposed
sensors for use in the quality assurance of these drugs. Results with a recovery of 99.5 % and
a mean standard deviations of ±1.3 % of the nominal are obtained which compared fairly well
with data obtained using the British pharmacopoeia method.