Title of article :
Displacement solid-phase extraction on mercapto-functionalized magnetite microspheres for inductively coupled plasma mass spectrometric determination of trace noble metals Original Research Article
Author/Authors :
Yan Li، نويسنده , , Yan-Feng Huang، نويسنده , , Yan Jiang، نويسنده , , Bo-lin Tian، نويسنده , , Fei Han ?، نويسنده , , Xiu-Ping Yan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
8
From page :
42
To page :
49
Abstract :
A flow injection online displacement solid-phase extraction (DSPE) via magnetic immobilization of mercapto-functionalized magnetite microspheres onto the inner walls of a knotted reactor (KR) coupled with inductively coupled plasma mass spectrometry was developed for selective preconcentration and determination of trace noble metals (Ru, Rh, Pd, Pt, Ir and Au) in complex matrices. Online DSPE of 2.7 mL aqueous solution gave the enhancement factors of 32–46 for the six noble metals in comparison with direct nebulization of aqueous sample solution, and the detection limits (3 s) of 2.1 ng L−1 for Ru, 1.9 ng L−1 for Rh, 2.5 ng L−1 for Pd, 1.8 ng L−1 for Ir, 1.9 ng L−1 for Pt and 1.7 ng L−1 for Au. The sample throughput of the developed method was about 20 samples h−1, and the relative standard deviation for eleven replicate determinations of the noble metals at the 30 ng L−1 level ranged from 1.2% to 2.1%. The recoveries of Ru, Rh, Pd, Pt, Ir and Au still maintained 90% even after successive 140 cycles of DSPE. The developed method was successfully applied to selective determination of trace Ru, Rh, Pd, Pt, Ir and Au in complex matrices.
Keywords :
Noble metals , Magnetic immobilization , Inductively coupled plasma mass spectrometry , Displacement solid-phase extraction
Journal title :
Analytica Chimica Acta
Serial Year :
2011
Journal title :
Analytica Chimica Acta
Record number :
1026404
Link To Document :
بازگشت