Title of article :
Matrix effects in inductively coupled plasma mass spectrometry: A review
Author/Authors :
Christian Agatemor، نويسنده , , Diane Beauchemin، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
18
From page :
66
To page :
83
Abstract :
Fundamental research on non-spectroscopic interferences, also known as matrix effects, in inductively coupled plasma (ICP) mass spectrometry with sample introduction using nebulization is critically and exclusively examined in this review, starting with fundamental processes that may be a source of matrix effects during sample introduction, ion generation in the ICP, ion extraction through the interface, and ion transport through the ion optics to the detector. Various methods for attenuating matrix effects are then reviewed and illustrated with some examples. Instead of exhaustively reviewing the literature, representative references are used to comprehensively describe the main issues, several of which are also common to ICP optical emission spectrometry.
Keywords :
Inductively coupled plasma mass spectrometry , Matrix effects , Robust plasma , Non-spectroscopic interferences
Journal title :
Analytica Chimica Acta
Serial Year :
2011
Journal title :
Analytica Chimica Acta
Record number :
1026729
Link To Document :
بازگشت