Title of article :
Improved detection of phosphopeptides by negative ion matrix-assisted laser desorption/ionization mass spectrometry using a proton sponge co-matrix Original Research Article
Author/Authors :
Shu Zhang، نويسنده , , Zhongping Yao، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
6
From page :
77
To page :
82
Abstract :
Analysis of phosphopeptides is an important task in proteomic studies. Matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) is a technique very commonly used for such a purpose. Analysis of phosphopeptides by MALDI-MS is, however, still a challenging task due to the low ionization efficiency of phosphopeptides. In this study, we reported that by using a proton sponge 1,8-bis(dimethyl-amino)naphthalene (DMAN) as a co-matrix, detection of phosphopeptides by negative ion MALDI-MS could be greatly improved. Combination of DMAN with another matrix 6-aza-2-thiothymine (ATT) and additive diammonium hydrogen citrate (DHC) allowed much lower limit of detection, significantly reduced signal suppression effects and improved position-to-position reproducibility for detection of phosphopeptides by negative ion MALDI-MS. Potential applications of the matrix system in qualification of phosphopeptides and analysis of proteolytic digests of phosphorylated proteins were also demonstrated in this study.
Keywords :
Proton sponge , Negative ion , Phosphopeptides , MALDI , Matrix , mass spectrometry