Title of article :
Determination of sulphide in effluent streams using a solid-phase lead(II) chromate reactor incorporated into a flow-injection system Original Research Article
Author/Authors :
J.F. van Staden، نويسنده , , L.G. Kluever، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
The determination of sulphide in waste water (effluent) from a dairy farm, using a solid-phase reactor incorporated into a flow-injection system, was investigated. The reactor packing material consisted of PbCrO4 suspended on alumina, and the chromate ions were released into the carrier stream when PbS, which is less soluble than PbCrO4, was formed. The chromate ions were detected spectrophotometrically at 370 nm. The linear response range of the method was between 10 and 50 mg l−1, with a detection limit of 3.7 mg l−1. The relative standard deviation was <3% for n=12 at 30 mg l−1 S2−.
Keywords :
Flow-injection , Chromate , Sulphide , Solid-phase reactor , Effluent
Journal title :
Analytica Chimica Acta