Title of article
Multivariate calibration of polycyclic aromatic hydrocarbon mixtures from excitation–emission fluorescence spectra Original Research Article
Author/Authors
J.L. Beltr?n، نويسنده , , R. Ferrer، نويسنده , , J. Guiteras، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
9
From page
311
To page
319
Abstract
The excitation–emission fluorescence spectra (EEM) of mixtures of 10 polycyclic aromatic hydrocarbons (PAHs) have been analyzed using different multivariate calibration procedures (partial least squares regression, PLSR; and parallel factor analysis, PARAFAC). The compounds studied were anthracene, benz[a]anthracene, benzo[a]pyrene, chrysene, phenanthrene, fluoranthene, fluorene, naphthalene, perylene and pyrene.
Several algorithms (PARAFAC, PLS1, PLS2, Tri-PLS1 and Tri-PLS2) were tested for the quantification of these compounds from the EEM spectra of a set of standards, at ng ml−1 levels. The best results were obtained by the application of a three-way multivariate calibration method (Tri-PLS2). This method was selected for the determination of samples of tap and mineral waters spiked with all these PAHs. Excitation and emission wavelength ranges were between 240 and 300 nm and 310 and 478 nm, respectively. In the working conditions, an EEM spectrum covering this range could be recorded in about 2 min.
Keywords
partial least squares , Multivariate calibration , PARAFAC , Polycyclic aromatic hydrocarbons , EEM fluorescence
Journal title
Analytica Chimica Acta
Serial Year
1998
Journal title
Analytica Chimica Acta
Record number
1027144
Link To Document