Title of article
Silver(I)-selective membrane electrode based on hexathia-18-crown-6 Original Research Article
Author/Authors
Mohammad Hossein Mashhadizadeh، نويسنده , , Mojtaba Shamsipur، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
6
From page
111
To page
116
Abstract
A PVC membrane electrode for Ag+ ions based on hexathia-18-crown-6 as membrane carrier was prepared. The sensor exhibits a Nernstian response for silver ions over a wide concentration range (3.2×10−3–6.0×10−6 M) with a slope of 59 mV decade−1. The limit of detection is 4.0×10−6 M (0.43 μg ml−1). It has a response time of <10 s and can be used for at least three months without any divergence in potential. The proposed electrode shows fairly a good discriminating ability towards Ag+ ion in comparison to some alkali, alkaline earth, transition and heavy metal ions. The electrode can be used in the pH range from 2.0 to 7.5. It was used as an indicator electrode in potentiometric titration of silver ion.
Keywords
Silver ion-selective electrode , Hexathia-18-crown-6 , Potentiometry , PVC membrane