Title of article
In situ preparation of a cholesterol biosensor: entrapment of cholesterol oxidase in an overoxidized polypyrrole film electrodeposited in a flow system: Determination of total cholesterol in serum Original Research Article
Author/Authors
Juan C. Vidal، نويسنده , , Esperanza Garcia، نويسنده , , Juan R. Castillo، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
10
From page
213
To page
222
Abstract
A straightforward method for constructing cholesterol biosensors by entrapment of cholesterol oxidase within a polypyrrole (PPy) film electropolymerized in a flow system is proposed. This method enables adjustment of the biosensor characteristics and features low reagent consumption. Prior to use, the PPy film is overoxidized to provide it with anion-exclusion properties in order to minimize the interference of electroactive species such as ascorbic acid and uric acid. A comprehensive preliminary study of the influence of 2-propanol and Triton X-100 added to the cholesterol solution on the flow-injection determination of this substrate, and the effects of different experimental variables on the biosensor response and its selectivity against interfering electroactive species was carried out. The proposed cholesterol oxidase based biosensor, named Pt/PPy–ChOx, was applied to the determination of cholesterol in reference serum samples, and the results were consistent with certified values.
Keywords
Polypyrrole electropolymerization , Cholesterol biosensor , Flow system , Cholesterol oxidase entrapment
Journal title
Analytica Chimica Acta
Serial Year
1999
Journal title
Analytica Chimica Acta
Record number
1027531
Link To Document