Title of article :
Measurement and control of hydrogen peroxide concentration in alkaline solution by means of amperometric sensor system Original Research Article
Author/Authors :
P Westbroek، نويسنده , , E Temmerman، نويسنده , , P Kiekens، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
6
From page :
423
To page :
428
Abstract :
An amperometric sensor system has been developed to measure and control the hydrogen peroxide concentration in a continuous and in-line way, over a wide concentration range (0.1–70 g/l), with high precision, accuracy and selectivity and over a long period of time without the need of recalibration or replacement of the active part of the sensor system. The system performs well in an industrial environment under the condition that the pH of the process liquid is at least 10.5.
Keywords :
Potentiometry , Conducting polymers , Electrodes , Organic acids , Capillary electrophoresis
Journal title :
Analytica Chimica Acta
Serial Year :
1999
Journal title :
Analytica Chimica Acta
Record number :
1027554
Link To Document :
بازگشت