Rapid and selective determination of peroxyacetic acid in disinfectants using flow injection analysis Original Research Article
Author/Authors :
Diedrich Harms، نويسنده , , Uwe Karst، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
6
From page :
233
To page :
238
Abstract :
The development and application of a flow injection analysis (FIA) method for the automated determination of peroxyacetic acids in the presence of hydrogen peroxide is described. The selective oxidation of 2,2′-Azino-bis(3-ethylbenzothiazoline)-6-sulphate) (ABTS) to a green radical cation (ABTS∗+) is the basis for this method. A reagent stream and a carrier stream into which the sample is injected are combined in a low dead volume mixing tee. A short reaction coil provides for a reaction time of 6 s after which detection is performed by UV/vis spectroscopically at λ=405 nm. Major advantages of the method are the simple experimental setup and the high selectivity towards hydrogen peroxide. Real samples have been investigated and the method has been validated using independent techniques including microplate spectrophotometry and liquid chromatography.
Keywords :
Flow injection analysis , Selective oxidation , Peroxyacetic acid