Title of article :
Effect of AC voltage frequency on the sensitivity of conductometric detection in microseparation techniques Original Research Article
Author/Authors :
Franti?ek Opekar، نويسنده , , Ivan Jel??nek، نويسنده , , Anton??n Troj?nek، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Pages :
6
From page :
101
To page :
106
Abstract :
Conductometric detectors employed in microseparation methods are designed to primarily match the separation criteria. Consequently, their geometry is seldom optimal from the point of view of conductometry. An unsuitable detection cell geometry is manifested in the dependence of the conductometric cell constant on the frequency of the AC signal applied to the electrodes. Variations in the cell constant value cause changes in the measured signal even if the solution conductivity is constant. By varying the frequency of the AC signal it is thus possible to find an optimal frequency at which the highest detection sensitivity is attained. The optimum frequency depends both on the cell geometry and on the conductivity of the mobile phase solution.
Keywords :
Detection , Conductometry , Microseparation methods