• Title of article

    Measurement of enantiomeric purity by ratio chromatograms with a photometric detector using quartz plates as a multiple retarder Original Research Article

  • Author/Authors

    Hiroyuki Nakazawa، نويسنده , , Tomoko Yamada، نويسنده , , Takaho Watanabe، نويسنده , , Atsushi Yamamoto ، نويسنده , , Akinobu Matsunaga، نويسنده , , Masayuki Nishimura، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    6
  • From page
    125
  • To page
    130
  • Abstract
    The measurement of enantiomeric purity is presented using the dual-wavelength detection function of a spectrophotometric detector. Between enantiomers, this method measures the difference in the absorbance ratio at adjacent 1/4 wavelengths that are produced by a thin quartz plate placed on a flow-cell. The use of a split-type flow-cell achieved high sensitivity. Moreover, the determination of absolute quantities of the analyte was successful from the peak shape by adjusting the ratio of effluent velocity between the reference and sample cells to 1/3. This method was applied to the determination of camphor in pharmaceutical preparations.
  • Keywords
    circular dichroism , HPLC , Enantiomeric purity , Retarder , Ratio chromatograms
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1999
  • Journal title
    Analytica Chimica Acta
  • Record number