Title of article
Determination of adenosine phosphate by co-electroosmotic capillary electrophoresis with amperometric detection Original Research Article
Author/Authors
Chonggang Fu، نويسنده , , Linan Song، نويسنده , , Yuzhi Fang، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
5
From page
259
To page
263
Abstract
A co-electroosmotic capillary electrophoretic (CE) system with amperometric detection (AD) was used for the simultaneous determination of adenosine triphosphate (ATP), adenosine diphosphate (ADP) and adenosine monophosphate (AMP). The optimum electrolyte consisted of 0.02 M phosphate adjusted to pH 5.40 and 0.15 mM cetyltrimethylammonium bromide (CTAB). Baseline separation was obtained within 21 min. Using a carbon fiber disc electrode at a potential of +1.20 V (vs. Ag/AgCl), a good linear relation between peak current and concentration for each analyte was obtained in the range of 5 ∼ 150 μg/ml. For the studied system, the precisions of migration time and peak current (in terms of relative standard deviation, n = 5) were in the range of 1.5 ∼ 1.8% and 2.0 ∼ 3.1%, respectively. The method was demonstrated by analyzing ATP, ADP and AMP in real medicinal tablets.
Keywords
Co-electroosmotic capillary electrophoresis , Amperometry , Adenosine phosphate