  • Abstract
    In this work, a microfluidic-chip based system for liquid-phase microextraction (LPME-chip) was developed. Sample solutions were pumped into the LPME-chip with a micro-syringe pump at a flow rate of 3–4 μL min−1. Inside the LPME chip, the sample was in direct contact with a supported liquid membrane (SLM) composed of 0.2 μL dodecyl acetate immobilized in the pores of a flat membrane of polypropylene (25 μm thickness). On the other side of the SLM, the acceptor phase was present. The acceptor phase was either pumped at 1 μL min−1 during extraction or kept stagnant (stop-flow). Amitriptyline, methadone, haloperidol, loperamide, and pethidine were selected as model analytes, and they were extracted from alkaline sample solution, through the SLM, and into 10 mM HCl or 100 mM HCOOH functioning as acceptor phase. Subsequently, the acceptor phase was either analyzed off-line by capillary electrophoresis for exact quantification, or on-line by UV detection or electrospray ionization mass spectrometry for time profiling of concentrations. The LPME-chip was found to be highly effective, and extraction efficiencies were in the range of 52–91%. When the flow of acceptor phase was turned off during extraction (stop-flow), analyte enrichment increased linearly with the extraction time. After 10 min as an example, amitriptyline was enriched by a factor of 42 from only 30 μL sample solution, and after 120 min amitriptyline was enriched by a factor of 500 from 320 μL sample solution. This suggested that the LPME-chip has great potentials for very efficient analyte enrichments from limited sample volumes in the future.
  • Keywords
    Liquid-phase microextraction , Microfluidic-chip , Biological fluids , Enrichment , Electrospray ionization mass spectrometry
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2012
  • Journal title
    Analytica Chimica Acta
  • Record number

    1028522