• Title of article

    Development of an automatic multi-channel ink-jet ejection chemiluminescence system and its application to the determination of horseradish peroxidase Original Research Article

  • Author/Authors

    Fengming Chen، نويسنده , , Zhen Lin، نويسنده , , Yongzan Zheng، نويسنده , , Hulie Zeng، نويسنده , , Hizuru Nakajima، نويسنده , , Katsumi Uchiyama ، نويسنده , , Jin-ming Lin، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    6
  • From page
    77
  • To page
    82
  • Abstract
    In this work, an automatic multi-channel ink-jet for chemiluminescence (CL) analysis was developed. The four-channel ink-jet device was controlled by a home-made circuit. Differing from the classic flow injection CL, the whole procedure for CL analysis was automatically completed on a hydrophobic glass side. CL reaction of luminal and hydrogen peroxide for the determination of horseradish peroxidase (HRP) was selected as an application to automatic CL analysis platform. All solutions delivered by different channels were precisely ejected to the same position of the glass slide for the CL analysis. The consumption of reaction solution was reduced to nanoliter level. The whole CL analysis could be completed in less than 4 min, which was benefited from the prompt solution mixing in small size of droplet. The CL intensity increased linearly with HRP concentration in the range from 0.01 to 0.5 μg mL−1. The limit of detection (LOD) (S/N = 3) was 0.005 μg mL−1. Finally, the automatic CL system could also be used for the detection of HRP in HRP–protein conjugates, which showed its practical application in immunoassay.
  • Keywords
    Horseradish peroxidase , Chemiluminescence , Ink-jet , Multi-channel device
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2012
  • Journal title
    Analytica Chimica Acta
  • Record number