• Title of article

    A new beryllium ion-selective membrane electrode based on dibenzo(perhydrotriazino)aza-14-crown-4 ether Original Research Article

  • Author/Authors

    Vinod Kumar Gupta، نويسنده , , Ashok Kumar Singh، نويسنده , , Naveen Mergu، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    7
  • From page
    44
  • To page
    50
  • Abstract
    Beryllium(II) selective electrodes have been fabricated based on poly(vinyl chloride) (PVC) matrix membranes containing newly synthesized neutral carrier dibenzo(perhydrotriazino)aza-14-crown-4 ethers as ionophore. Best performance was exhibited by the membrane having a composition ionophore (IIa):PVC:sodium tetraphenylborate (NaTPB):tributyl phosphate (TBP) in the ratio (w/w; mg) of 5:30:3:65. This membrane worked well over a wide concentration range 7.6 × 10−6 to 1.0 × 10−1 M of Be2+ with a Nernstian slope of 30.7 mV per decade of beryllium activity. The response time of the sensor is 15 s and the membrane can be used over a period of 4 months with good reproducibility. The proposed electrode works well in a wide pH range 3.0–9.0. It was successfully applied to the determination of beryllium in a mineral sample.
  • Keywords
    Beryllium selective electrode , Dibenzo(perhydrotriazino)aza-14-crown-4 ether , Potentiometry , Poly(vinyl chloride) membrane
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2012
  • Journal title
    Analytica Chimica Acta
  • Record number