Title of article
Ultrasensitive one-step rapid detection of ochratoxin A by the folding-based electrochemical aptasensor Original Research Article
Author/Authors
Jingjing Wu، نويسنده , , Huaqin Chu، نويسنده , , Zhanlong Mei، نويسنده , , Yi Deng، نويسنده , , Feng Xue، نويسنده , , Lei Zheng، نويسنده , , Wei Chen، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
5
From page
27
To page
31
Abstract
A one-step electrochemical aptasensor using the thiol- and methylene blue- (MB-) dual-labeled aptamer modified gold electrode for determination of ochratoxin A (OTA) was presented in this research. The aptamer against OTA was covalently immobilized on the surface of the electrode by the self-assembly effect and used as recognition probes for OTA detection by the binding induced folding of the aptamer. Under the optimal conditions, the developed electrochemical aptasensor demonstrated a wide linear range from 0.1 pg mL−1 to 1000 pg mL−1 with the limit of detection (LOD) of 0.095 pg mL−1, which was an extraordinary sensitivity compared with other common methods for OTA detection. Moreover, as a practical application, this proposed electrochemical aptasensor was used to monitor the OTA level in red wine samples without any special pretreatment and with satisfactory results obtained. Study results showed that this electrochemical aptasensor could be a potential useful platform for on-site OTA measurement in real complex samples.
Keywords
Aptamer , Electrochemical aptasensor , One-step detection , Ochratoxin A detection , Ultrasensitive
Journal title
Analytica Chimica Acta
Serial Year
2012
Journal title
Analytica Chimica Acta
Record number
1028840
Link To Document