  • Abstract
    Dissolved sulfide ions are separated and detected by using ion chromatography (IC) and electrochemical detection with a glassy carbon electrode modified by electrochemical deposition of palladium particles. The influence of various experimental factors such as pH, loading of palladium, applied potential and flow rate of mobile phase is evaluated. The effect of some common organic and inorganic compounds on the amperometric signal is also evaluated. Under optimal chromatographic conditions, the detection limit for sulfide (S/N = 3) was 0.3 μM (15 pmol injected) and the calibration plot was linear for 0.5–350 μM sulfide (r = 0.9964, n = 16). The separation and detection of sulfides in human blood serum is presented as an example of an application to real specimens.
  • Keywords
    Ion chromatography , Flow injection analysis , Electrochemical detection , Sulfide measurements