Title of article :
Selective stopped-flow injection spectrophotometric determination of palladium(II) in hydrogenation and automobile exhaust gas converter catalysts Original Research Article
Author/Authors :
Aristidis N. Anthemidis، نويسنده , , Demetrius G. Themelis، نويسنده , , John A Stratis، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
7
From page :
161
To page :
167
Abstract :
A stopped-flow injection spectrophotometric method is reported for the determination of palladium(II), using 2,2′-dipyridyl-2-pyridylhydrazone (DPPH) as a colour forming reagent. The absorbance of the Pd(II)-DPPH complex was monitored at 540 nm, at pH≈0.3. The various chemical and physical parameters were optimized and a study of interfering ions was also carried out. The calibration graph has two linear parts, first in the range 0–25 mg l−1 (sr=0.44%, r=0.9999) with a detection limit of cL=0.084 mg l−1 and the second in the range 25–60 mg l−1 (sr=1.6%, r=0.9996). In both cases the sampling rate was 30 injections per hour. The method is very selective because the strongly acidic medium used prevented the formation of complexes of the reagent with other ions. The method was successfully applied to the determination of palladium in hydrogenation and automobile exhaust gas converters catalysts. The relative standard deviation of the mean values and the recovery ranged between 0.6 and 1.6% and 97.0–102.6%, respectively.
Keywords :
Palladium , Spectrophotometry , Stopped-flow injection , Hydrogenation and automobile catalysts
Journal title :
Analytica Chimica Acta
Serial Year :
2000
Journal title :
Analytica Chimica Acta
Record number :
1029080
Link To Document :
بازگشت