Title of article
Lead ion-selective membrane electrode based on 1,10-dibenzyl-1,10-diaza-18-crown-6 Original Research Article
Author/Authors
M.F. Mousavi، نويسنده , , S Sahari، نويسنده , , N Alizadeh، نويسنده , , M Shamsipur، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
6
From page
189
To page
194
Abstract
A PVC membrane lead(II) ion-selective electrode has been constructed using 1,10-dibenzyl-1,10-diaza-18-crown-6 (DBzDA18C6) as membrane carrier. The sensor shows a Nernstian response for Pb(II) ions over a wide concentration range (5.0×10−5 to 1.0×10−2 M). It has a fast response time and can be used for more than 10 months without any considerable divergence in potentials. The proposed sensor exhibits comparatively good selectivities with respect to alkali, alkaline earth and some transition and heavy metal ions. It was employed as an indicator electrode in potentiometric titration of EDTA with lead ions and used for direct determination of lead in waste water.
Keywords
Ion-selective membrane electrode , Potentiometry , Crown ether , lead
Journal title
Analytica Chimica Acta
Serial Year
2000
Journal title
Analytica Chimica Acta
Record number
1029176
Link To Document