• Title of article

    Lead ion-selective membrane electrode based on 1,10-dibenzyl-1,10-diaza-18-crown-6 Original Research Article

  • Author/Authors

    M.F. Mousavi، نويسنده , , S Sahari، نويسنده , , N Alizadeh، نويسنده , , M Shamsipur، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    6
  • From page
    189
  • To page
    194
  • Abstract
    A PVC membrane lead(II) ion-selective electrode has been constructed using 1,10-dibenzyl-1,10-diaza-18-crown-6 (DBzDA18C6) as membrane carrier. The sensor shows a Nernstian response for Pb(II) ions over a wide concentration range (5.0×10−5 to 1.0×10−2 M). It has a fast response time and can be used for more than 10 months without any considerable divergence in potentials. The proposed sensor exhibits comparatively good selectivities with respect to alkali, alkaline earth and some transition and heavy metal ions. It was employed as an indicator electrode in potentiometric titration of EDTA with lead ions and used for direct determination of lead in waste water.
  • Keywords
    Ion-selective membrane electrode , Potentiometry , Crown ether , lead
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2000
  • Journal title
    Analytica Chimica Acta
  • Record number