• Title of article

    Improved porous silicon microarray based prostate specific antigen immunoassay by optimized surface density of the capture antibody Original Research Article

  • Author/Authors

    Sangwook Lee، نويسنده , , Soyoun Kim، نويسنده , , Johan Malm، نويسنده , , Ok Chan Jeong، نويسنده , , Hans Lilja، نويسنده , , Thomas Laurell، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    7
  • From page
    108
  • To page
    114
  • Abstract
    Enriching the surface density of immobilized capture antibodies enhances the detection signal of antibody sandwich microarrays. In this study, we improved the detection sensitivity of our previously developed P-Si (porous silicon) antibody microarray by optimizing concentrations of the capturing antibody. We investigated immunoassays using a P-Si microarray at three different capture antibody (PSA – prostate specific antigen) concentrations, analyzing the influence of the antibody density on the assay detection sensitivity. The LOD (limit of detection) for PSA was 2.5 ng mL−1, 80 pg mL−1, and 800 fg mL−1 when arraying the PSA antibody, H117 at the concentration 15 μg mL−1, 35 μg mL−1, and 154 μg mL−1, respectively. We further investigated PSA spiked into human female serum in the range of 800 fg mL−1 to 500 ng mL−1. The microarray showed a LOD of 800 fg mL−1 and a dynamic range of 800 fg mL−1 to 80 ng mL−1 in serum spiked samples.
  • Keywords
    Prostate specific antigen , Porous silicon , Antibody microarray , Sandwich immunoassay
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2013
  • Journal title
    Analytica Chimica Acta
  • Record number