• Title of article

    On-line determination of iron(II) and iron(III) using a spectrophotometric sequential injection system Original Research Article

  • Author/Authors

    Ludwig V. Mulaudzi، نويسنده , , Jacobus F. van Staden، نويسنده , , Raluca I. Stefan، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    15
  • From page
    35
  • To page
    49
  • Abstract
    The application of sequential injection analysis (SIA) for metal speciation analysis is presented. The proposed method was used for the quantitative discrimination of two iron species, Fe(II) and Fe(III). Tiron was used as the chromogenic reagent for Fe(III) and total iron after Fe(II) was oxidised to Fe(III) by H2O2. The linear range is from 0.15 to 100.00 and 0.30 to 80.00 mg l−1 for Fe(III) and Fe(II), respectively, with a %R.S.D. of 1.3 and 0.8% for Fe(III) and Fe(II), respectively. The detection limit (DL) was found to be 0.10 and 0.15 mg l−1 for Fe(III) and Fe(II), respectively. The system is fully computerised and able to process 30 samples h−1.
  • Keywords
    Speciation , Sequential injection analysis , Iron(II) , Iron(III)
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2002
  • Journal title
    Analytica Chimica Acta
  • Record number